The following parameters can be selected and, if required, changed:
Value Offset (reduction of zero point 0-5 ppm),
Output (Output of actual value 0 / 4-20 mA),
Alarm 1 Limit Value (Limit Alarm 1, 0-15 ppm),
Alarm 2 Limit Value (Limit Alarm 2, 0-15 ppm),
Alarm 1 Time Delay (Delay of response for alarm 1, 0-15 sec),
Alarm 2 Time Delay (Delay of response for alarm 2, 0-15 sec),
Datalog Export (transferring the storage data to a PC by means of PC software).
1. Press key
P to enter the programming mode. The amber
system LED will be illuminated and the window showing
the parameter that was changed last appears.
2. Press
 or  to scroll forwards / backwards until the
desired parameter is shown.
3. Press key
P (P = change) to activate the parameter chan-
ge. To quit the programming mode without any changes
press key OK (OK = back). The window that was active
before you switched the mode will reappear.
4. After having activated the parameter change by pressing
key P use the keys +/ to adjust the value within the
permissible limits.
5. To cancel the new setting press key
P (P = abort). The old
value will then remain valid.
6. To confirm and activate the new value press key OK
(OK = confirm). The parameter window appears again.
7. Press key OK (OK = back) to quit the programming mode.
The window that was active before you switched the mode
will reappear.
Programming mode
Commissioning Procedure – continued –
Parameter
act Value Offset
Value : 0 ppm
= prev = next
= backOKP
= change
Parameter
act Value Offset
Value : 2 ppm
= inc Para = dec Para
= confirmOKP
+
= abort
Parameter
act Value Offset
Value : 2 ppm
= prev = next
= backOKP
= change